ISSN 1004-4140
    CN 11-3017/P

    超高分辨力CT腕关节成像适宜曝光参数的选择

    Selection of Appropriate Exposure Parameters for Wrist Joint Imaging Using Ultra-high-resolution CT

    • 摘要: 目的:探讨Ultra3D-SE 超高分辨率 CT 扫描仪(U-HRCT)腕关节扫描中的适宜曝光参数。方法:以自制弹道明胶与成人手腕骨制成的腕关节模体作为被照体,应用U-HRCT及多层螺旋CT(MSCT)对模体进行曝光:U-HRCT管电压在60~100 kV范围内间隔5 kV变化,管电流在2~10 mA范围内间隔2 mA变化,固定曝光时间对模体进行扫描获取CT图像。MSCT在120 kV及自动管电流调制(ATCM)技术下,对腕关节模体进行常规CT扫描。用热释光剂量计(TLD)对上述两种扫描方案每次曝光时腕关节模体的辐射剂量(取两片测量的平均值)进行测量,并计算品质因数(FOM)。对不同条件下的信噪比(SNR)、对比噪声比(CNR)、背景噪声(SD)、剂量面积乘积(DAP)、入射面空气比释动能(Ka,e)、品质因数(FOM)进行分析,确定最佳的曝光参数。结果:曝光参数选择60 kV和2 mA时,CNR、SNR、DAP和(Ka,e)值最小(分别为3.5、5.4、23.68 mGy·cm2和0.40 mGy),曝光条件选择100 kV和10 mA时,CNR、SNR、DAP和(Ka,e)最大(分别为13.8、18.1、368.75 mGy·cm2和12.21 mGy)。60 kV和2 mA时SD最大为383.8,100 kV和10 mA时SD最小为64.7。FOM在70 kV和2 mA时最大为84.3。70 kV和2 mA组的(Ka,e)为0.73 mGy,MSCT的(Ka,e)为5.48 mGy,经计算U-HRCT的(Ka,e)比MSCT降低约86.68%。结论:管电压和管电流的选择对U-HRCT图像质量与辐射剂量影响较大,70 kV和2 mA既能满足临床诊断需求,又能降低受检者的辐射剂量,可作为U-HRCT的推荐扫描参数。

       

      Abstract: Objective: To investigate the optimal exposure parameters for ultra-high-resolution computed tomography (U-HRCT) in wrist joint imaging. Methods: A wrist joint phantom composed of self-made ballistic gelatin and adult wrist bones was used as the imaging object. The phantom was scanned using both U-HRCT and multi-slice spiral CT (MSCT). For U-HRCT, the tube voltage was varied from 60 to 100 kV in 5 kV increments, while the tube current ranged from 2 to 10 mA in 2 mA increments, with a fixed exposure time for image acquisition. For MSCT, routine wrist CT scanning was performed at 120 kV using automatic tube current modulation (ATCM). Thermoluminescent dosimeters (TLD) were used to measure the radiation dose delivered to the wrist phantom during each exposure under both scanning protocols, and the average value from two TLD measurements was recorded. The figure of merit (FOM) was subsequently calculated. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), background noise (SD), dose–area product (DAP), entrance surface air kerma (Ka,e), and FOM under different exposure conditions were analyzed to determine the optimal exposure parameters. Results: At an exposure setting of 60 kV and 2 mA, the CNR, SNR, DAP, and Ka,e values were the lowest, measuring 3.5, 5.4, 23.68 mGy·cm², and 0.40 mGy, respectively. At 100 kV and 10 mA, the CNR, SNR, DAP, and Ka,e values reached their maximum values of 13.8, 18.1, 368.75 mGy·cm², and 12.21 mGy, respectively. The SD was highest at 383.8 under 60 kV and 2 mA, whereas it was lowest at 64.7 under 100 kV and 10 mA. The maximum FOM value of 84.3 was achieved at 70 kV and 2 mA. Under the 70 kV and 2 mA condition, the Ka,e value was 0.73 mGy, compared with 5.48 mGy for MSCT. Calculations showed that the Ka,e of U-HRCT was reduced by approximately 86.68% relative to MSCT. Conclusion: Tube voltage and tube current significantly affect both image quality and radiation dose in U-HRCT imaging. The exposure setting of 70 kV and 2 mA can satisfy clinical diagnostic requirements while substantially reducing radiation dose to the subject, and is therefore recommended as the optimal scanning protocol for U-HRCT.

       

    /

    返回文章
    返回