ISSN 1004-4140
CN 11-3017/P
程英, 关计添, 金阳, 郭岳霖, 张远芳. 64层螺旋CT冠状动脉低剂量成像初探[J]. CT理论与应用研究, 2009, 18(3): 91-95.
引用本文: 程英, 关计添, 金阳, 郭岳霖, 张远芳. 64层螺旋CT冠状动脉低剂量成像初探[J]. CT理论与应用研究, 2009, 18(3): 91-95.
CHENG Ying, GUAN Ji-tian, JIN Yang, GUO Yue-lin, ZHANG Yuan-fang. 64-MSCT Coronary Angiography with Reduction Radiation Dose[J]. CT Theory and Applications, 2009, 18(3): 91-95.
Citation: CHENG Ying, GUAN Ji-tian, JIN Yang, GUO Yue-lin, ZHANG Yuan-fang. 64-MSCT Coronary Angiography with Reduction Radiation Dose[J]. CT Theory and Applications, 2009, 18(3): 91-95.

64层螺旋CT冠状动脉低剂量成像初探

64-MSCT Coronary Angiography with Reduction Radiation Dose

  • 摘要: 目的:优化扫描模式,降低冠状动脉成像辐射剂量。方法:60名患者随机分为两组,每组30人。钙化积分分别应用430mA和80mA管电流扫描;延迟时间测定使用团注测定和智能跟踪;冠状动脉成像包括心电门控非管电流调节和管电流调节模式,分别计算所有组平均ED和图像SNR。结果:钙化积分平均ED从2.35±0.42降至0.42±0.03;延迟时间平均ED从1.39±0.14降至0.13±0.04;两组冠状动脉成像平均ED从28.22±4.90降至21.14±1.54。两组钙化积分平均SNR组间比较有显著性差异,而两组冠状动脉成像模式平均SNR组间比较无显著性差异。结论:应用80mA扫描钙化积分、智能跟踪测试延迟时间和心电门控调节管电流模式进行冠状动脉成像,可显著降低辐射剂量并获得诊断需要的图像。

     

    Abstract: Objective: To optimize the scan protocol,and reduce the radiation dose in coronary angiography with 64-MSCT.Methods: 60 patients were randomly divided into two groups,30 included in each group.Calcium score images were scanned by tube current of 430 mA and 80 mA;delayed time was detected by timing bolus and Smart Prep,respectively;Coronary angiography was performed using ECG-gated protocol without tube current modulation and that with tube current modulation.In all groups,the mean effective dose and signal noise of ratio were calculated and compared.Results: Mean ED of calcium score scan obviously decreased from 2.35 ±0.42 of 430 mA to 0.42 ±0.03 of 80 mA;that of the delayed time detection reduce increasingly from 1.39 ±0.14 by timing bolus to 0.13 ±0.04 by Smart Prep;and that of coronary angiography protocol reduced from 28.22 ±4.90 without tube current modulation to 21.14 ±1.54 with tube current modulation.There were significant differences between mean SNR by 430 mA with that by 80 mA in calcium score scans,but there were no significant differences between mean SNR without tube current modulation with that with tube current modulation.Conclusion: Using the scan protocol with tube current of 80 mA in calcium score,delayed time detection of Smart Prep,and ECG-gated tube current modulation,the radiation dose of coronary angiography with 64 MSCT could be reduced increasingly,and image quality achieves the diagnostic desire.

     

/

返回文章
返回