ISSN 1004-4140
CN 11-3017/P
JIANG D, QIN L X. Spectral Filtration Sn 150 kV Combined with Advanced Simulated Iterative Reconstruction in Lumbar Computed Tomography Examination[J]. CT Theory and Applications, 2024, 33(1): 49-55. DOI: 10.15953/j.ctta.2022.240. (in Chinese).
Citation: JIANG D, QIN L X. Spectral Filtration Sn 150 kV Combined with Advanced Simulated Iterative Reconstruction in Lumbar Computed Tomography Examination[J]. CT Theory and Applications, 2024, 33(1): 49-55. DOI: 10.15953/j.ctta.2022.240. (in Chinese).

Spectral Filtration Sn 150 kV Combined with Advanced Simulated Iterative Reconstruction in Lumbar Computed Tomography Examination

More Information
  • Received Date: November 29, 2022
  • Revised Date: April 01, 2023
  • Accepted Date: April 05, 2023
  • Available Online: September 06, 2023
  • Objective: To study the feasibility of energy spectrum purification Sn 150 kV combined with Advanced Simulated Iterative Reconstruction (ADMIRE) in computed tomography (CT) examination of the lumbar spine. Methods: A total of 88 patients aged 25-65 years with body mass indexes (BMI) between 18.5~25 kg/m2 were randomly divided into a control group (group A) and an experimental group (group B), with 44 cases in each group. The conventional tube voltage, used for the control group, was 120 kV, while the experimental tube voltage was Sn 150 kV. All other imaging parameters were consistent. After the inspection, the image quality and radiation dose of the two groups were compared. Results: The objective evaluation of noise in group A and group B were statistically different, with the noise of group A higher than that of group B. The ICC of the three diagnosticians in the subjective evaluation was 0.769, indicating a good consistency of evaluation. The radiation dose of group B was 30.31% lower than that of group A. Conclusion: The combination of energy spectrum purification Sn 150 kV and ADMIRE can not only effectively reduce the radiation dose, but can also guarantee a high-quality image, which can be widely used in adult lumbar CT.

  • [1]
    石清磊, 赵红梅, 张玲, 等. 自适应统计迭代重建算法对腹部CT扫描中图像质量和辐射剂量影响的模体研究[J]. 中华放射学杂志, 2013,47(4): 326−329. DOI: 10.3760/cma.j.issn.1005-1201.2013.04.008.

    SHI Q L, ZHAO H M, ZHANG L, et al. Model study on the effect of adaptive statistical iterative reconstruction algorithm on image quality and radiation dose in abdominal CT scanning[J]. The Chinese Journal of Radiology, 2013, 47(4): 326−329. DOI: 10.3760/cma.j.issn.1005-1201.2013.04.008. (in Chinese).
    [2]
    YANG C H, WU T H, CHIOU Y Y, et al. Imaging quality and diagnostic reliability of low-dose computed tomography lumbar spine for evaluating patients with spinal disorders[J]. The Spine Jourmal, 2014, 14(11): 2682−2690. DOI: 10.1016/j.spinee.2014.03.007.
    [3]
    ALBRECHT M, NANCE J W, SCHOEPF U, et al. Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol[J]. European Radiology, 2018, 28(5): 2134-2142. DOI:10.1007/s00330-017-5150-z.
    [4]
    陈楚韩, 綦维维, 刘晓怡, 等. 深度学习重建算法结合低管电压技术提高常规腹部增强CT图像小血管显示水平的价值[J]. 中华放射学杂志, 2022,56(11): 1168−1174. DOI: 10.3760/cma.j.cn112149-20220218-00131.

    CHEN C H, QI W W, LIU X Y, et al. Value of deep learning reconstruction algorithm combined with low tube voltage technology to improve the display level of small vessels in conventional abdominal enhanced CT images[J]. The Chinese Journal of Radiology, 2022, 56(11): 1168−1174. DOI: 10.3760/cma.j.cn112149-20220218-00131. (in Chinese).
    [5]
    姬宇虹. 螺旋CT扫描结合iDose4技术在肺结节中的临床应用[J]. 实用放射学杂志, 2018,(34): 126−127. DOI: 10.3969/j.issn.1002-1671.2018.01.034.

    JI Y H. Clinical application of spiral CT scanning combined with iDose4 technique in pulmonary nodules[J]. Practical Radiology Journal, 2018, (34): 126−127. DOI: 10.3969/j.issn.1002-1671.2018.01.034. (in Chinese).
    [6]
    李翔, 王翔, 杨欣, 等. 自动管电流技术联合人工智能在COVID-19 肺部低剂量CT中的应用价值[J]. 放射学实践, 2020,(35): 1082−1086. DOI: 10.13609/j.cnki.1000-0313.2018.10.017.

    LI X, WANG X, YANG X, et al. Application value of automatic tube current technology combined with artificial intelligence in low-dose lung CT of COVID-19[J]. Radiology Practices, 2020, (35): 1082−1086. DOI: 10.13609/j.cnki.1000-0313.2018.10.017. (in Chinese).
    [7]
    唐慧, 杨军乐, 贺太平, 等. 3种不同迭代重建技术在胸部低剂量CT扫描中的应用[J]. 实用放射学杂志, 2022,(6): 1003−1006. DOI: 10.3969/j.issn.1002-1671.2022.06.033.

    TANG H, YANG J L, HE T P, et al. Application of three different iterative reconstruction techniques in low dose chest CT scanning[J]. Journal of Practical Radiology, 2022, (6): 1003−1006. DOI: 10.3969/j.issn.1002-1671.2022.06.033. (in Chinese).
    [8]
    于安宏, 杨雅朋, 马延贺, 等. 超低剂量CT能谱纯化技术联合高级建模迭代重建对显示肺磨玻璃结节及测量其体积的影响: 体模研究[J]. 中国医学影像技术, 2022,(38): 765−769. DOI: 10.13929/j.issn.1003-3289.2022.05.030.

    YU A H, YANG Y P, MA Y H, et al. Effect of ultra-low dose CT energy spectrum purification technique combined with advanced modeling iterative reconstruction on displaying and measuring lung ground glass nodules: A body model study[J]. Chinese Journal of Medical Imaging, 2022, (38): 765−769. DOI: 10.13929/j.issn.1003-3289.2022.05.030. (in Chinese).
    [9]
    王青云, 梁俊生, 江世健, 等. 第三代双源CT能谱纯化技术在鼻窦CT扫描中的应用[J]. 实用放射学杂志, 2021,37(7): 1192−1195. DOI: 10.3969/j.issn.1002-1671.2021.07.034.

    WANG Q Y, LIANG J S, JIANG S J, et al. Application of the third generation dual-source CT energy spectrum purification technology in sinus CT scanning[J]. Journal of Applied Radiology, 2021, 37(7): 1192−1195. DOI: 10.3969/j.issn.1002-1671.2021.07.034. (in Chinese).
    [10]
    何付权, 李亚飞, 罗倩, 等. 双源CT Sn 100 kV能谱纯化技术在胸部CT筛查中的应用[J]. 实用放射学杂志, 2021,37(5): 846−849, 856. DOI: 10.3969/j.issn.846-849.2021.05.036.

    HE F Q, LI Y F, LUO Q, et al. Application of dual-source CT Sn 100 kV energy spectrum purification technology in chest CT screening[J]. Journal of Applied Radiology, 2021, 37(5): 846−849, 856. DOI: 10.3969/j.issn.846-849.2021.05.036. (in Chinese).
    [11]
    姜一, 田葵, 沙晋璐, 等. 基于能谱纯化结合迭代重建算法在继发性肺结核患者低剂量CT检查中的应用[J]. CT理论与应用研究, 2022,31(1): 95−101. DOI: 10.15953/j.1004-4140.2022.31.01.11.

    JIANG Y, TIAN K, SHA J L, et al. Application of energy spectrum purification combined with iterative reconstruction algorithm in low-dose CT examination of patients with secondary tuberculosis[J]. CT Theory and Applications, 2022, 31(1): 95−101. DOI: 10.15953/j.1004-4140.2022.31.01.11. (in Chinese).
    [12]
    蒋东, 秦立新, 李晟, 等. 能谱纯化结合高级迭代重建在儿童胸部超低剂量CT的应用初探[J]. 实用放射学杂志, 2021,(37): 1536−1538. DOI: 10.3969/j.issn.1002-1671.2021.09.033.

    JIANG D, QIN L X, LI S, et al. Application of energy spectrum purification combined with advanced iterative reconstruction in pediatric chest ultra-low dose CT[J]. Journal of Applied Radiology, 2021, (37): 1536−1538. DOI: 10.3969/j.issn.1002-1671.2021.09.033. (in Chinese).
    [13]
    袁子龙, 刘涛, 张彪, 等. 能谱纯化联合器官剂量调制技术对婴幼儿头部CT扫描器官剂量及图像质量影响的体模研究[J]. 中华放射医学与防护杂志, 2022,42(3): 225−229. DOI: 10.3760/cma.j.cn112271-20211008-00405.

    YUAN Z L, LIU T, ZHANG B, et al. Study on the effect of energy spectrum purification combined with organ dose modulation technology on the organ dose and image quality of infant head CT scan[J]. The Chinese Journal of Radiological Medicine and Protection, 2022, 42(3): 225−229. DOI: 10.3760/cma.j.cn112271-20211008-00405. (in Chinese).
    [14]
    沈梓璇, 王晓野, 卢晓娟. 腰椎检查中运用螺旋CT低剂量扫描对患者辐射剂量以及图像评分的影响分析[J]. 影像研究与医学应用, 2021,5(22): 108−109. DOI: 10.3969/j.issn.2096-3807.2021.22.050.

    SHEN Z X, WANG X Y, LU X J. Effect of low dose spiral CT scan on radiation dose and image score of patients in lumbar examination[J]. Imaging Research and Medical Application, 2021, 5(22): 108−109. DOI: 10.3969/j.issn.2096-3807.2021.22.050. (in Chinese).
    [15]
    王帅, 华晨辰, 朱宗明, 等. 能谱纯化(Sn 150 kV)扫描在全腹部CT的应用价值[J]. 影像研究与医学应用, 2021,5(17): 14−16. DOI: 10.3969/j.issn.2096-3807.2021.17.007.

    WANG S, HUA C C, ZHU Z M, et al. Application of energy spectrum purification (Sn 150 kV) in total abdominal CT[J]. Medical Imaging Research and Application, 2021, 5(17): 14−16. DOI: 10.3969/j.issn.2096-3807.2021.17.007. (in Chinese).
    [16]
    顾海峰, 鲍雪琴, 王清清, 等. 高级建模迭代重建对冠状动脉钙化积分的影响[J]. 放射学实践, 2022,5(8): 1028−1034. DOI: 10.13609/j.cnki.1000-0313.2022.08.019.

    GU H F, BAO X Q, WANG Q Q, et al. Senior modeling iterative reconstruction effect on coronary artery calcium score[J]. Journal of Radiology Practice, 2022, 5(8): 1028−1034. DOI: 10.13609/j.cnki.1000-0313.2022.08.019. (in Chinese).
    [17]
    SCHLUNK S, BYRAM B. Combining ADMIRE and MV to improve image quality[J]. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2022: 2651-2662. DOI:10.1109/TUFFC.2022.3194548.
  • Cited by

    Periodical cited type(1)

    1. 赖小珍,邓玲玲. 比较MRI与CT检查对肝脏孤立性结节鉴别的诊断价值. 现代诊断与治疗. 2024(13): 1976-1977+1980 .

    Other cited types(0)

Catalog

    Article views (151) PDF downloads (22) Cited by(1)
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return