Effect of External Foreign Bodies on Patient Radiation Dose in Chest CT

More Information
  • Received Date: March 20, 2025
  • Revised Date: April 10, 2025
  • Accepted Date: April 14, 2025
  • Available Online: April 29, 2025
  • To investigate the effect of external foreign bodies on radiation dose during chest computed tomography (CT) scans under automatic tube current-modulation technology and to assess the necessity of re-scanning a scout image, we retrospectively categorize historical data into foreign body and non-foreign body cohorts. Propensity score matching is employed to balance five covariates: patient gender, examination age, number of image layers, scanning-bed height, and average cross-sectional area of patients. After balancing, we statistically analyze the effect of foreign bodies on the volume CT dose index (CTDIvol) and dose-length product (DLP), and compare the dose increment with the scout-image dose. The study includes 17342 cases, with 4740 cases (27.3%) in the treatment group. Before matching, all five covariates in the treatment group are significantly lower than those in the control group. After matching, 3944 pairs of balanced samples are obtained and no significant differences are observed between the two groups in the five covariates. The treatment group shows slightly higher CTDIvol and DLP values compared with the control group. External foreign bodies can slightly increase the radiation dose, although the incremental dose does not differ statistically from that of the scout image. Operators should flexibly determine whether to re-scan the scout image based on specific clinical circumstances.

  • [1]
    TSALAFOUTAS I A, ALKHAZZAM S, ALNAEMI H, et al. Evaluation of automatic tube current modulation of CT scanners using a dedicated and the CTDI dosimetry phantoms[J]. Journal of Applied Clinical Medical Physics, 2022, 23(7): e13620. DOI: 10.1002/acm2.13620.
    [2]
    吴德红, 陈少波, 付传明, 等. ATCM技术在CTU中降低辐射剂量的应用探讨[J]. CT理论与应用研究, 2015, 24(2): 319-325. DOI: 10.15953/j.1004-4140.2015.24.02.18.

    WU D H, CHEN S B, FU C M, et al. Value of automatic tube current modulation techniques to reducing the radiation dose in CTU[J]. CT Theory and Applications, 2015, 24(2): 319-325. DOI: 10.15953/j.1004-4140.2015.24.02.18. (in Chinese).
    [3]
    AUSTIN P C. A critical appraisal of propensity-score matching in the medical literature between 1996 and 2003[J]. Statistics in Medicine, 2008, 27(12): 2037-2049. DOI: 10.1002/sim.3150.
    [4]
    ROSENBAUM P R, RUBIN D B. Constructing a control group using multivariate matched sampling methods that incorporate the propensity score[J]. The American Statistician, 1985, 39(1): 33-38. DOI: 10.1080/00031305.1985.10479383.
    [5]
    冉姗姗, 綦维维, 张淼, 等. 多参数设置对低剂量胸部CT扫描图像质量及辐射剂量的影响[J]. 中国医学影像技术, 2018, 34(1): 113-117. DOI: 10.13929/j.1003-3289.201704039.

    RAN S S, QI W W, ZHANG M, et al. Impact of multiple parameters setting on image quality and radiation dose in low-dose chest CT scanning[J]. Chinese Journal of Medical Imaging Technology, 2018, 34(1): 113-117. DOI: 10.13929/j.1003-3289.201704039. (in Chinese).
    [6]
    刘丹丹, 崔莹, 赵波, 等. 定位像扫描参数对胸部CT影像质量和辐射剂量影响的模体研究[J]. 中华放射医学与防护杂志, 2021, 41(3): 217-221. DOI: 10.3760/cma.j.issn.0254-5098.2021.03.011.

    LIU D D, CUI Y, ZHAO B, et al. The influence of scout scanning parameters on image quality and radiation dose of chest CT: A phantom study[J]. Chinese Journal of Radiological Medicine and Protection, 2021, 41(3): 217-221. DOI: 10.3760/cma.j.issn.0254-5098.2021.03.011. (in Chinese).
    [7]
    杨政君, 张昂, 陈勇, 等. 辐射剂量和管电压对CT图像质量的影响: 基于任务的图像质量评价[J]. CT理论与应用研究, 2022, 31(2): 211-217. DOI: 10.15953/j.ctta.2021.060.

    YANG Z J, ZHANG A, CHEN Y, et al. The effect of radiation dose and tube potential on image quality of CT: A task-based image quality assessment[J]. CT Theory and Applications, 2022, 31(2): 211-217. DOI: 10.15953/j.ctta.2021.060. (in Chinese).
    [8]
    李海梅, 刘康, 隋岩, 等. 不同CT扫描条件对胸部模体实性结节人工智能检出效率及辐射剂量的影响[J]. 中华放射医学与防护杂志, 2023, 43(3): 216-221. DOI: 10.3760/cma.j.cn112271-20221114-00446.

    LI H M, LIU K, SUI Y, et al. The influence of different CT scanning protocols on AI detection efficiency and radiation dose of solid nodules in chest phantom[J]. Chinese Journal of Radiological Medicine and Protection, 2023, 43(3): 216-221. DOI: 10.3760/cma.j.cn112271-20221114-00446. (in Chinese).
    [9]
    张宇, 何长久, 杨梦犀, 等. 自动管电压联合自动管电流在冠状动脉CT血管成像中的应用[J]. 中国医疗设备, 2023, 38(5): 115-119. DOI: 10.3969/j.issn.1674-1633.2023.05.021.

    ZHANG Y, HE C J, YANG M X, et al. Application of automatic tube voltage combined with automatic tube current techniques in coronary CT angiography[J]. China Medical Devices, 2023, 38(5): 115-119. DOI: 10.3969/j.issn.1674-1633.2023.05.021. (in Chinese).
    [10]
    邵培育, 苏小蓉. 自动管电流调制技术对腹部CT图像质量及辐射剂量的影响[J]. 医疗装备, 2022, 35(22): 32-34. DOI: 10.3969/j.issn.1002-2376.2022.22.010.
    [11]
    罗锐. 探讨双源CT自动管电流调制技术在冠状动脉成像中降低辐射剂量的价值[J]. 影像研究与医学应用, 2020, 4(3): 17-18.

    LUO R. To explore the value of dual-source CT automatic tube current modulation technique in reducing radiation dose in coronary artery imaging[J]. Journal of Imaging Research and Medical Applications, 2020, 4(3): 17-18. (in Chinese).
    [12]
    李翔, 王翔, 杨欣, 等. 自动管电流技术联合人工智能在COVID-19肺部低剂量CT中的应用价值[J]. 放射学实践, 2020, 35(9): 1082-1086. DOI: 10.13609/j.cnki.1000-0313.2020.09.001.

    LI X, WANG X, YANG X, et al. Application of automatic tube current technique combined with artificial intelligence in the low dose CT for COVID-19[J]. Radiologic Practice, 2020, 35(9): 1082-1086. DOI: 10.13609/j.cnki.1000-0313.2020.09.001. (in Chinese).
    [13]
    黄涛, 马隆佰, 陆玉敏, 等. 80kV管电压下迭代重建算法联合自动管电流调制技术在低剂量头颈部CT血管造影中的应用[J]. 广西医学, 2020, 42(18): 2364-2368. DOI: 10.11675/j.issn.0253-4304.2020.18.09.

    HUANG T, MA L B, LU Y M, et al. Application of iterative reconstruction algorithm combined with automatic tube current modulation technology to low-dose computed tomography angiography of head and neck in the setting of tube voltage of 80kV[J]. Guangxi Medical Journal, 2020, 42(18): 2364-2368. DOI: 10.11675/j.issn.0253-4304.2020.18.09. (in Chinese).

Catalog

    Article views (60) PDF downloads (6) Cited by()