Estimation of Energy Spectra of CT System And Beam-Hardening Correction Algorithm
-
1. Department of Mathematics, Northwest University, ChinaXi’an 710069;
2. Division of Applied Mathematics, Beijing Information Technology Institute, China Beijing 100101
More Information
-
-
- Received Date:
April 17, 2003
- Available Online:
December 28, 2022
-
Abstract
The authors present an algorithm which can estimate, by means of a standard cylinder or wedge model, the energy spectra of the CT system and the mass attenuation coefficient of the model at the same time. An algorithm on beam-hardening correction of CT image for single material object is presented as well. The simulation results show that the algorithm for estimation of the energy spectra can give a reasonable estimation of the energy spectra, and that beam-hardening correction algorithm is more precise than polynomial approximation correction algorithm and more rapid than common iterative algorithm.
-
-
-