Malfunction Servicing of Philips T5-NT MRI Set

More Information
  • Received Date: August 19, 2001
  • Available Online: December 28, 2022
  • Half black screen was shown in Philips T5-NT MRI Scanner and with its message. This trouble was associated with environment temperature, checking machine and find the bad parts TC-420 board at the best power up time. The malfunction disappeared after this board was replaced.