ISSN 1004-4140
    CN 11-3017/P
    Liu Hong, Liu Ying. Malfunction Servicing of Philips T5-NT MRI Set[J]. CT Theory and Applications, 2001, 10(4): 31-32.
    Citation: Liu Hong, Liu Ying. Malfunction Servicing of Philips T5-NT MRI Set[J]. CT Theory and Applications, 2001, 10(4): 31-32.

    Malfunction Servicing of Philips T5-NT MRI Set

    • Half black screen was shown in Philips T5-NT MRI Scanner and with its message. This trouble was associated with environment temperature, checking machine and find the bad parts TC-420 board at the best power up time. The malfunction disappeared after this board was replaced.
    • loading

    Catalog

      /

      DownLoad:  Full-Size Img  PowerPoint