Liu Hong, Liu Ying. Malfunction Servicing of Philips T5-NT MRI Set[J]. CT Theory and Applications, 2001, 10(4): 31-32.
Citation: Liu Hong, Liu Ying. Malfunction Servicing of Philips T5-NT MRI Set[J]. CT Theory and Applications, 2001, 10(4): 31-32.

Malfunction Servicing of Philips T5-NT MRI Set

  • Half black screen was shown in Philips T5-NT MRI Scanner and with its message. This trouble was associated with environment temperature, checking machine and find the bad parts TC-420 board at the best power up time. The malfunction disappeared after this board was replaced.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return