Liu Hong, Liu Ying. Detecting A Soft Trouble of Philips MRI RF Amplifier[J]. CT Theory and Applications, 2001, 10(3): 39-41.
Citation: Liu Hong, Liu Ying. Detecting A Soft Trouble of Philips MRI RF Amplifier[J]. CT Theory and Applications, 2001, 10(3): 39-41.

Detecting A Soft Trouble of Philips MRI RF Amplifier

  • A soft trouble of Philips MRI RF amplifier occurred with message -'Bias current out of range' on operating console screen, and the scan stopped. After checking the RF amplifier by 'Test Software', and adjusting maximum RF power, the malfunction unit-RF power amplifier board was found, and it was proved by replaced the board.