Xu Weiqing. The Analysis of Ring Artifacts SOMATOM AR[J]. CT Theory and Applications, 2001, 10(1): 36-40.
Citation: Xu Weiqing. The Analysis of Ring Artifacts SOMATOM AR[J]. CT Theory and Applications, 2001, 10(1): 36-40.

The Analysis of Ring Artifacts SOMATOM AR

More Information
  • Received Date: December 19, 2000
  • Available Online: December 28, 2022
  • The reason why ring artifacts appear in Somatom AR is analysed. The appearance of ring artifacts is related with working temperature and humidity. The hardware (includes Detector, hybrid and A/D inverter) multifunction is the main causes. The ring artifacts is also caused by position and angle's dis-adjustment between DAS, Tube and Collimator.
  • Related Articles

    [1]LIU Bicheng, YI Xi, ZONG Chunguang, XU Yanwei, LI Liang. Ring-artifact Correction Method for Large-size Object CT Images Based on Gradient Featured Cluster Analysis[J]. CT Theory and Applications, 2024, 33(6): 781-789. DOI: 10.15953/j.ctta.2024.153
    [2]LIU Xiaojing, WU Hongying, MA Jinqiang, LEI Ziqiao, YU Jianming. Application of Dual-layer Detector Spectral Computed Tomography Virtual Non-contrast Scanning in Enhanced Chest Computed Tomography for Children[J]. CT Theory and Applications, 2024, 33(6): 709-716. DOI: 10.15953/j.ctta.2024.123
    [3]NING xianying, HU Liu, FAN Wenliang, YU Jianming, LEI Ziqiao. The Application Value of Monoenergetic Imaging on Dual-layer Spectral Detector Computed Tomography in Dacryocystography[J]. CT Theory and Applications, 2024, 33(6): 677-682. DOI: 10.15953/j.ctta.2024.144
    [4]XU Pin, CHEN Shi, YUAN Gang, ZHANG Yin, ZHENG Jian, XING Xiao-man, SUN Ming-shan. Study of Spectral CT Based on a Photon Counting Detector[J]. CT Theory and Applications, 2015, 24(6): 819-825. DOI: 10.15953/j.1004-4140.2015.24.06.07
    [5]LI Zhong-chuan, MENG Fan-yong, YANG Min. The Correction of Fan-beam CT Image Artifacts Induced by Deflection of Linear-detectors[J]. CT Theory and Applications, 2014, 23(2): 237-248.
    [6]ZHANG Hua, HUANG Kui-dong, SHI Yi-kai, LI Ming-jun. Ring Artifact Correction Method Based on Air Scan for Cone-beam CT[J]. CT Theory and Applications, 2012, 21(2): 247-254.
    [7]HU Dong-cai, CHEN Hao, ZHANG Ding-hua. Scatter Correction Method for Flat-Panel Detector-Based Cone Beam CT[J]. CT Theory and Applications, 2009, 18(1): 16-22.
    [8]DING Wei-hua, ZHANG Jian-jun, CHEN Hou-qun, LIU Shao-cong. Test Equipment Problems on Studying Concrete Using X-Ray CT[J]. CT Theory and Applications, 2007, 16(2): 42-47.
    [9]WANG Yuan, XU Zhou, CHEN Hao, LI Ming, JIN Xiao, HU He-ping, GAN Kong-yin, ZHANG Cheng-xin, LU He-ping. Data acquiring System for High Energy Industrial CT Based on Flat-Panel Detector[J]. CT Theory and Applications, 2006, 15(3): 53-56.
    [10]WANG Xue-li. Medical CT Artifact Analysis: Reason and Correction[J]. CT Theory and Applications, 2004, 13(3): 1-8.
  • Cited by

    Periodical cited type(5)

    1. 刘素萍,魏玲玲. 针对性护理干预在CT增强扫描中的应用效果研究. 现代医药卫生. 2022(04): 681-684 .
    2. 吴建峰,李林静,方春,王豪杰,祝峰. 个体化碘对比剂注射速率方案在肝脏增强CT中提高增强质量一致性的价值. 中国医师杂志. 2022(03): 456-458 .
    3. 龚剑英,谢军,喻科,颜得新. 基于身体指数测量计算肝脏CT增强扫描个体化碘对比剂方案的研究. 中国CT和MRI杂志. 2022(10): 84-86 .
    4. 董相宇,方挺松,袁健祥,梁振华,李均洪. 基于BSA及固定碘速率探讨不同浓度对比剂对肝静脉CT成像效果的影响. CT理论与应用研究. 2021(04): 466-476 . 本站查看
    5. 孟晶晶,韩耀华,孙大明. 对比剂浓度和患者门静脉CT影像质量的关系分析. 河北医学. 2021(09): 1486-1490 .

    Other cited types(0)

Catalog

    Article views (1575) PDF downloads (10) Cited by(5)
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return