WANG Rui-qiong, FU Xia, DENG Jian-guo, LIU Jian-hua, CAI Li, QUAN Xiang, CHEN Xia, CAO Shen, SUN Jun. Coronary Computed Tomography Angiography Using Low-kV and Tube Current Modulation in Subjects with Normal Body Mass Indices: A Feasibility Study[J]. CT Theory and Applications, 2013, 22(2): 345-352.
Citation: WANG Rui-qiong, FU Xia, DENG Jian-guo, LIU Jian-hua, CAI Li, QUAN Xiang, CHEN Xia, CAO Shen, SUN Jun. Coronary Computed Tomography Angiography Using Low-kV and Tube Current Modulation in Subjects with Normal Body Mass Indices: A Feasibility Study[J]. CT Theory and Applications, 2013, 22(2): 345-352.

Coronary Computed Tomography Angiography Using Low-kV and Tube Current Modulation in Subjects with Normal Body Mass Indices: A Feasibility Study

More Information
  • Received Date: December 26, 2012
  • Available Online: December 12, 2022
  • Objectlve: To assess the image quality and radiation dose of 80 kV electrocardiography(ECG)-gated CCTA in subjects with a normal body mass index(BMI),compared to 120 kV ECG-gated CCTA.Method: The CCTA images of 140 subjects with BMIs < 25 kg/m2 were analyzed.Seventy subjects underwent 120 kV CCTA,and the other 70 subjects underwent 80kV CCTA.Two blinded observers independently evaluated the subjective image quality of the coronary arteries.The objective image quality(signal-to-noise ratio(SNR) and contrast-to-noise ratio(CNR)) and radiation dose were also measured in each group.Results: Although the objective image quality of the 80 kV protocol images was significantly poorer than that of 120 kV protocol images(P < 0.05),there was no significant difference in the subjective image quality between the two groups(P > 0.05).The radiation dose was reduced by 12mSv with the 80 kV protocol and by 82% with the 80 kV and ECG-based tube current modulation than one with the 120 kV protocol(P < 0.05).Conclusion: The low tube voltage CCTA protocol using 80 kV allows significant reduction of the radiation dose without impairing the subjective image quality in subjects with normal BMIs.

Catalog

    Article views PDF downloads Cited by()
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return