CHANG Xin, WANG Xu, FENG Chen-lu. The Effect of Axial and Spiral Scan Mode in Emergency Head CT on Image Quality and Radiation Dosage[J]. CT Theory and Applications, 2015, 24(1): 117-122. DOI: 10.15953/j.1004-4140.2015.24.01.14
Citation: CHANG Xin, WANG Xu, FENG Chen-lu. The Effect of Axial and Spiral Scan Mode in Emergency Head CT on Image Quality and Radiation Dosage[J]. CT Theory and Applications, 2015, 24(1): 117-122. DOI: 10.15953/j.1004-4140.2015.24.01.14

The Effect of Axial and Spiral Scan Mode in Emergency Head CT on Image Quality and Radiation Dosage

More Information
  • Received Date: October 09, 2014
  • Available Online: December 05, 2022
  • Objective: to study the effect of axial and spiral scan mode in head CT on image quality and radiation dosage, and provide the evidence for emergency patient to choice the reasonable scanning mode. Methods: A total of 20 patients were taken the emergency CT with spiral scan mode and with axial scan mode in the first recheck. Scan parameters: 120 k V, 290 m A, 0.8 s/rol, slice thickness 5 mm, scanning interval 5 mm, pitch 0.969:1, standard reconstruction algorithm. To measure the CT value and background noise of the white matter and grey matter and calculate the SNR and CNR, and evaluate image noise, artifacts of the skull base, gray matter-white matter differentiation. After scanning, the radiation dosage was obtained from the dose report. Paired sample t test and chi-square test were used to compare the difference of image quality between two scanning modes. Results: Using axial scanning mode, the CT value of grey matter and white matter was(38.02 ±2.23)HU and(27.01 ±1.72)HU respectively, and using spiral scanning mode, the CT value of grey matter and white matter was(37.76 ±2.22) HU and(26.99 ±1.69) HU respectively. There is no significant difference in CT value, SNR, CNR between two modes(P > 0.05, t = 0.045, 0.119, 0.502). There is no significant difference in image noise, artifacts of the skull base, the scores of grey matter-white matter differentiation between two modes(P > 0.05, χ2= 3.142, 0.0, 3.159). Effective doses was 66 m Sv in spiral mode, and 43 m Sv in axial mode, but the scanning time of spiral mode is shorter, and the reconstruction is helpful to the diagnosis. Conclusions: In term of both image quality and radiation dosage, the axial scanning mode is the preferred choice in the emergency head CT, but spiral scanning mode should be taken for the restless and non-cooperative patients.
  • Related Articles

    [1]HU Yanjun, WEI Xuan, ZHANG Yu, TIAN Xin, WANG Likun, JIANG Guilian, NIU Yantao. Effect of External Foreign Bodies on Patient Radiation Dose in Chest CT[J]. CT Theory and Applications. DOI: 10.15953/j.ctta.2025.103
    [2]QI Rongfei, JIANG Wenlong, HAO Yunyun, YANG Liusha, CHANG Yingjuan, WU Zhibin. A Study of Image Quality and Radiation Dose in Lower Extremity Computed Tomography Angiography Using Caudo-cranial Flash Scanning[J]. CT Theory and Applications, 2025, 34(3): 385-391. DOI: 10.15953/j.ctta.2024.210
    [3]LIU Kang, YANG Dewu, ZHANG yongxian, LIU dandan, GAI xinghui, CHEN zhaorui, GONG hao, SUI Yan. Impact of Different Scanning Modes and Noise Index on Inner Ear CT Image Quality and Radiation Dose[J]. CT Theory and Applications. DOI: 10.15953/j.ctta.2024.230
    [4]WANG Chensi, HONG Yao, FANG Huimin, DING Xuelu, YANG Yifan, TIAN Xiangbao, CHEN Wenjing, WANG Yongsheng. Based on the Effect of Different Pitches on Neck CTA Radiation Dose and Image Quality under Automatic Tube Current Modulation Technology[J]. CT Theory and Applications, 2024, 33(3): 309-315. DOI: 10.15953/j.ctta.2023.148
    [5]YANG Zhengjun, ZHANG Ang, CHEN Yong, JIANG Jiang, WANG Lingyun, ZHANG Yong, ZHANG Xuan, QI Xiaofeng. The Effect of Radiation Dose and Tube Potential on Image Quality of CT: A Task-based Image Quality Assessment[J]. CT Theory and Applications, 2022, 31(2): 211-217. DOI: 10.15953/j.ctta.2021.060
    [6]CAO Yu, ZHANG Yong-xian. Comparison of Multi-slice Spiral CT Imaging for Cardiac Coronary Arteries with Different Radiation Doses[J]. CT Theory and Applications, 2016, 25(1): 95-101. DOI: 10.15953/j.1004-4140.2016.25.01.12
    [7]WU De-hong, CHEN Shao-bo, FU Chuan-ming, XU Lin, CHEN Lun-gang, YU Guang-ju. Value of Automatic Tube Current Modulation Techniques to Reducing the Radiation Dose in CTU[J]. CT Theory and Applications, 2015, 24(2): 319-325. DOI: 10.15953/j.1004-4140.2015.24.02.18
    [8]BAO Li-wei, WANG Ze-feng, SUN Zhen-ting, MENG Ling-xin, ZHAO Lei. Evaluation of Image Quality and Radiation Dose for Low-tube Voltage Head and Neck CTA[J]. CT Theory and Applications, 2015, 24(1): 87-94. DOI: 10.15953/j.1004-4140.2015.24.01.10
    [9]FU Chuan-ming, WU De-hong, XU Lin, CHEN Lun-gang, WANG Zhong-ping, XIONG Yin. Reduce the Radiation Dose in the Lower Extremities to Explore the Feasibility of CTA[J]. CT Theory and Applications, 2014, 23(3): 467-473.
    [10]ZHANG Hai-bo, FU Chuan-ming, CHEN Lun-gang, XU Lin, WANG Kai-hua. Pitch Optimizing Combination of Thoracic and Abdominal CTA Image Quality and Radiation Dose Discussion[J]. CT Theory and Applications, 2012, 21(2): 305-311.

Catalog

    Article views (903) PDF downloads (12) Cited by()
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return