ISSN 1004-4140
CN 11-3017/P
FU Chuanming, CHEN Lungang, CHEN Wen, XU Lin, WANG Sirong. Research on the Contrast Medium's Best Injection Combined with Low Tube Voltage in Aortic CTA[J]. CT Theory and Applications, 2019, 28(4): 447-454. DOI: 10.15953/j.1004-4140.2019.28.04.05
Citation: FU Chuanming, CHEN Lungang, CHEN Wen, XU Lin, WANG Sirong. Research on the Contrast Medium's Best Injection Combined with Low Tube Voltage in Aortic CTA[J]. CT Theory and Applications, 2019, 28(4): 447-454. DOI: 10.15953/j.1004-4140.2019.28.04.05

Research on the Contrast Medium's Best Injection Combined with Low Tube Voltage in Aortic CTA

More Information
  • Received Date: April 09, 2019
  • Available Online: November 07, 2021
  • Published Date: August 24, 2019
  • Objective: To explore the feasibility of contrast medium's best injection combined with low tube voltage in aortic CTA.Methods: Divided 80 patients who will be done aortic CTA examination into A, B two groups(n=40) by different contrast medium concentrations and tube voltages, while group A is divided into group A1, group A2(n=20) by different contrast medium concentrations and injection rate.Group A1: Injecting 302 mg I/mL Iohexol as contract medium, injection rate 4 mL/s, tube voltages 100 kV;group A2: Injecting 300 mg I/mL Iohexol as contract medium, injection rate 3.5 mL/s, tube voltages 100 kV;group B: Injecting 350 mg I/mL Iohexol as contract medium, injection rate 3.5 mL/s, tube vantage 120 kV;Other scanning parameters: Injection contrast dose, reconstruction mode, automatic tube current regulation technologies are consistent.Two independent sample tests are used(Mann-Whitney U test for non-normal distribution) to compare the clinic data, radiation dose of CT, subjective and objective evaluation of image quality in the two groups.Results: There was no significant difference in age, gender, body mass index(BMI) between group A and group B(P>0.05);there was also no statistical significance in the subjective and objective evaluation of image quality(P>0.05).The CNR of group B was better than group A, which met the diagnostic requirements, the CT values of the aortic trunk and branches were all>300 HU.There was statistically significant differences between group A and group B in radiation dose and iodine content(P<0.01).Effective radiation dose(ED) of group A was 17.73%lower than that of group B, and iodine content was 14.29%lower.Conclusion: injection of 300 mgI/mL Iohexol solution and injection rate of 3.5 mL/s combined with low tube voltage(100 kV) obtained satisfactory image quality in aortic CTA examination, providing accurate diagnostic basis for clinical practice, which not only reduced the radiation dose, but also reduced the iodine content.
  • Related Articles

    [1]WANG Qian, SHI Shikui. Morphological Characteristics of Intracranial Aneurysms and Applicability of CT Angiography in Risk Assessment of Rupture[J]. CT Theory and Applications, 2024, 33(4): 503-510. DOI: 10.15953/j.ctta.2023.144
    [2]FENG Mengya, LI Bing, WU Min, HUA Ting, TANG Guangyu. Application of New Iterative Reconstruction Algorithm Based on Artificial Intelligence and Low Tube Voltage in Reducing the Radiation Dose of Coronary CTA in Obese Patients[J]. CT Theory and Applications, 2023, 32(6): 777-782. DOI: 10.15953/j.ctta.2022.148
    [3]WANG Yongsheng, ZHANG Pengyu, LI Fangzhong, WANG Zhenghua, LU Mengyun, CHEN Lei, HE Junlin, CHEN Wenjing. Study on the Influence of Head and Neck CTA Image Quality under Different Tube Voltages[J]. CT Theory and Applications, 2022, 31(5): 631-638. DOI: 10.15953/j.ctta.2022.026
    [4]FANG Shu, CHEN Linyu, CHEN Yong, DONG Haipeng, WANG Lan, MIN Jihua. Phantom Study on the Influence of CT Model and Tube Voltage on Image Quality[J]. CT Theory and Applications, 2022, 31(3): 345-350. DOI: 10.15953/j.ctta.2021.042
    [5]FU Chuan-ming, CHEN Shao-bo, WU De-hong, WANG Zhong-ping, XU Lin, CHEN Lun-gang, YU Guang-ju. The Application of Low Tube Voltage and Low Volume Contrast Agent Protocol in Aorta CTA[J]. CT Theory and Applications, 2015, 24(3): 437-444. DOI: 10.15953/j.1004-4140.2015.24.03.14
    [6]BAI Shao-jun, LI Li. The Diagnostic Value of Conventional CTA and Subtraction CTA on Carotid Artery Stenosis[J]. CT Theory and Applications, 2015, 24(3): 415-420. DOI: 10.15953/j.1004-4140.2015.24.03.11
    [7]BAO Li-wei, WANG Ze-feng, SUN Zhen-ting, MENG Ling-xin, ZHAO Lei. Evaluation of Image Quality and Radiation Dose for Low-tube Voltage Head and Neck CTA[J]. CT Theory and Applications, 2015, 24(1): 87-94. DOI: 10.15953/j.1004-4140.2015.24.01.10
    [8]CHEN Hong-cai, LU Jin-guo, PENG Wan-hong, XI Ren-gang. Diagnosis of Coronary Luminal Stenosis in Dual-source CT Coronary Angiography Using Low-tube-voltage[J]. CT Theory and Applications, 2014, 23(3): 371-377.
    [9]TIAN Shu-ping, YANG Li, WANG Zhan-yu, WANG Shou-hai, WANG Zi-jun, ZHANG Yan-qun. The Application of “Magic Glass” Function in Coronary CTA[J]. CT Theory and Applications, 2012, 21(4): 721-726.
    [10]LI Ruo-mei, WANG Si-wei, JIN Dan-dan. Low-dose CT Coronary Angiography with Status Quo of Tube Current and Voltage Parameters Optimization[J]. CT Theory and Applications, 2012, 21(3): 515-520.
  • Cited by

    Periodical cited type(7)

    1. 李明昕,刘杰,陈岩,高剑波,李国文,许晨思. 60kV联合深度学习重建算法在主动脉CT血管造影中的可行性研究. 影像研究与医学应用. 2024(06): 62-64 .
    2. 覃永平,许梅海,冯冬梅,杜芳,覃静,覃琳琳,申炜. 左右肘静脉注射对比剂对主动脉CT血管成像效果的对比研究. 中国CT和MRI杂志. 2023(05): 154-156 .
    3. 覃永平,许梅海. 左右肘静脉注射对比剂对不同部位CT血管成像效果影响的研究进展. 中国医药科学. 2022(01): 43-46+66 .
    4. 花庆兰. 对比剂不同流率推注方案对头颈部CTA图像清晰度、血管及周围结构对比度的影响研究. 世界复合医学. 2022(08): 10-13 .
    5. 黄萌,范小萍,严高武,刘玲. 低管电压和低碘浓度溶剂联合自适应统计迭代重建V技术在主动脉CT造影扫描中的可行性研究. 中国医学装备. 2022(12): 46-50 .
    6. 王娟,邓国莉,陆杰. 触发阈值为30 HU联合多期双流注射45 mL对比剂对大体重人群行头颈部联合CTA扫描的分析. 世界复合医学. 2022(12): 56-58+69 .
    7. 丘文科,李伟钦,梁树生. 64排CT头颈CTA检查中时间密度曲线法延时时间的优化. 中外医学研究. 2021(23): 92-95 .

    Other cited types(0)

Catalog

    Article views (278) PDF downloads (9) Cited by(7)
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return